Skip to main content

Table 1 Hg/Au electrode sensitivities (calibration slopes) measured in artificial seawater

From: How to overcome inter-electrode variability and instability to quantify dissolved oxygen, Fe(II), mn(II), and S(−II) in undisturbed soils and sediments using voltammetry

 

Sensitivity (10-2 nA/μM)

 

Mn(II)

Fe(II)

S(-II)

Artificial seawater (no organic carbon)

E1

7.4(1), 5.6(1)

2.71(7)

31.6(3), 30.0(3)

E2

7.7(1), 6.5(1)

3.11(8)

31(1), 23.1(2)

E3

6.9(1), 12.5(3)

5.3(1)

24.7(5), 55.3(4)

Artificial seawater with 0.85 (mg C)/L Pacific Ocean fulvic acid

E1

7.7(2)

3.26(9)

26.2(4)

E2

6.3(2)

2.11(6)

17(1)

E3

6.9(2)

2.66(8)

21.9(8)

Artificial seawater with 26 (mg C)/L Pony Lake fulvic acid

E1

-

5.4(1)

46(2)

E2

-

2.44(5)

21.59(5)

E3

-

2.77(6)

25.0(3)

  1. Figures in parentheses indicate the 90% confidence interval in the last digit.