Skip to main content

Table 2 Comparison of calibration slope ratios for Mn(II) and Fe(II) for different electrodes

From: How to overcome inter-electrode variability and instability to quantify dissolved oxygen, Fe(II), mn(II), and S(−II) in undisturbed soils and sediments using voltammetry

Date

Electrode

sMn

sFe

K = sMn/sFe

sS(-II)

K = sMn/sS(-II)

i(nA)

sOxygen

K = sMn/sOxygen

8/19/10

1

0.05607

0.0271

2.07

0.300

0.187

29.3

0.211

0.266

 

2

0.06931

0.0311

2.23

0.231

0.300

31.6

0.227

0.305

 

3

0.1246

0.0537

2.32

0.553

0.225

45

0.324

0.385

8/20/10

1

0.077

0.0326

2.36

0.262

0.294

27.8

0.200

0.385

 

2

0.0633

0.0211

3.00

0.168

0.377

20.2

0.145

0.436

 

3

0.06991

0.0266

2.63

0.219

0.319

23.7

0.171

0.410

average (all K values)

2.43

 

0.284

  

0.364

standard deviation

0.33

 

0.068

  

0.065

95% confidence interval

0.27

 

0.054

  

0.052

C.I./average

   

11%

 

19%

  

14%